VEGASON 51P 53P 53
Setup
Save the adjustments with " and you
are again in the menu window "
Meas. envi-
ronment
.
Meas. environment/Meas. conditions
Click in the menu window "
to " .
In the menu window "
Measuring conditions
you click on the options corresponding to
your application.
Confirm with .
After a few seconds, during which the adjust-
ments are permanently saved in the sensor,
you are again in the window "
Meas. environ-
ment
.
Meas. conditions/Sonic velocity
In the menu item "
Sonic velocity
adjustments
are only necessary when using the sensor in
a gas composition deviating from air. If you
measure in a gas composition deviating from
air, e.g. pure nitrogen, the sonic velocity
changes. Without correction, the measuring
result would be incorrect.
In the menu "
Sonic velocity
the sonic velocity
can be set to a value corresponding to a
different gas composition.
With " you save the adjustments in the
sensor.
Click in the window "
Meas. environment
to
" .
You are again in the menu window "
Sensor
optimisation
.